Dosimetric impact of uncorrected systematic yaw rotation in VMAT for peripheral lung SABR
Sarah Barrett1, Pierre Thirion2, Dean Harper2
1Applied Radiation Therapy Trinity, Discipline of Radiation Therapy, Trinity College Dublin, Ireland.
Aim:
This study aimed to evaluate the dosimetric impact of uncorrected yaw rotational error on both target coverage and OAR dose metrics in this patient population.
Background:
Rotational set up errors can be difficult to correct in lung VMAT SABR treatments, and may lead to a change in planned dose distributions.
Materials And Methods:
We retrospectively applied systematic yaw rotational errors in 1° degree increments up to -5° and +5° degrees in 16 VMAT SABR plans. The impact on PTV and OARs (oesophagus, spinal canal, heart, airway, chest wall, brachial plexus, lung) was evaluated using a variety of dose metrics. Changes were assessed in relation to percentage deviation from approved planned dose at 0 degrees.
Results:
Target coverage was largely unaffected with the largest mean and maximum percentage difference being 1.4% and 6% respectively to PTV D98% at +5 degrees yaw.Impact on OARs was varied. Minimal impact was observed in oesophagus, spinal canal, chest wall or lung dose metrics. Larger variations were observed in the heart, airway and brachial plexus. The largest mean and maximum percentage differences being 20.77% and 311% respectively at -5 degrees yaw to airway D0.1cc, however, the clinical impact was negligible as these variations were observed in metrics with minimal initial doses.
Conclusions:
No clinically unacceptable changes to dose metrics were observed in this patient cohort but large percentage deviations from approved dose metrics in OARs were noted. OARs with associated PRV structures appear more robust to uncorrected rotational error.
More Related Videos
08:30Intraperitoneal Glucose Tolerance Test, Measurement of Lung Function, and Fixation of the Lung to Study the Impact of Obesity and Impaired Metabolism on Pulmonary Outcomes
Published on: March 15, 2018
05:22Author Spotlight: Demonstrating Systematic Endobronchial Ultrasound to New Endoscopists
Published on: August 11, 2023
Related Concept Videos
07:28Donor Posterior Atrial Flap Rotation for Left Atrial Cuff Reconstruction in Lung Transplantation
08:30Intraperitoneal Glucose Tolerance Test, Measurement of Lung Function, and Fixation of the Lung to Study the Impact of Obesity and Impaired Metabolism on Pulmonary Outcomes
05:22Systematic Endobronchial Ultrasound - The Six Landmarks Approach
Random and Systematic Errors
10:03Uncoupling Coriolis Force and Rotating Buoyancy Effects on Full-Field Heat Transfer Properties of a Rotating Channel
Systematic Sampling Method
Systematic sampling is one of the simplest methods...