Background:
Aim of this study is to investigate the cyclic fatigue resistance of the Gold treated WaveOne Gold and the Firewire treated EdgeOne Fire instruments. The null hypotesis was that there were no differences between the lifespan of Gold treated and FireWire treated instruments when subjected to cyclic fatigue tests.
Material And Methods:
40 new NiTi instruments with a length of 25 mm were tested: 20 Wave One Gold Medium (WOG), tip size 35 and variable taper (Dentsply Maillefer, Ballaigues, Switzerland) and 20 Edge One Fire (EOF) (EdgeEndo, Albuquerque, New Mexico) tip size 35 and the same variable taper. A mobile support for the electric handpiece and a stainless-steel block containing the artificial canals were used. The same artificial root canal with a 90 degrees angle of curvature and 2 mm radius of curvature was used for all the tested instruments and the WOG counter-clock wise reciprocating motion with an engaging angle of 150° and a disengaging angle of 30° at 300 rpm, was selected for the test. All instruments were inserted at the same length (18mm) and then rotated in the same reciprocating motion until fracture occurred: the time was stopped as soon as the fracture was visible and video-recorded with a 1/100 sec chronometer. Differences among groups were statistically evaluated with an analysis of variance test ANOVA (significance level was set at p<0.05).
Results:
Mean values of time to fracture (TtF) for EOF instruments were 28,00 seconds (SD +/- 2,64) and for WOG instruments were 14,67 seconds (SD +/- 2,41). Statistical analysis found significant differences between the TtF of the two instruments (p<0,05).
Conclusions:
Firewire instruments resulted to be about two times more resistant to cyclic fatigue when compared with identical instruments made with Gold treatment. Key words:Endodontics, NiTi, Waveone Gold, EdgeOne Fire, Cyclic Fatigue.
Related Concept Videos
Fatigue of Metals
The importance of studying metal fatigue in civil infrastructure projects was brought into the spotlight by the collapse of the Silver Bridge in Point Pleasant, West Virginia in 1967. The eyebar-chain suspension bridge over the Ohio River collapsed during evening rush hour, killing 46 people as a result of the failure of a single eyebar with a small 0.1-inch defect. The defect reached a...
07:14Guided Endodontics: Three-Dimensional Planning and Template-Aided Preparation of Endodontic Access Cavities
09:48Construction and Operation of a Light-driven Gold Nanorod Rotary Motor System
03:28Lifespan Protocol for Drosophila melanogaster: Generating Survival Curves to Identify Differences in Fly Longevity
09:25The Treadmill Fatigue Test: A Simple, High-throughput Assay of Fatigue-like Behavior for the Mouse
Cyclic Voltammetry (CV)
A Cyclic Voltammetry (CV) experiment involves the scan of a range of potential voltages while measuring current. In the CV experiment, the potential of an immersed, stationary electrode is scanned from a predetermined starting potential to a final value (called the switching potential) and then the reverse scan is obtained. This gives a 'cyclic' sweep of potentials and the current vs. potential curve derived from...