Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
Simple and cost-effective determination of polychlorinated biphenyls in insulating oils using an ionic liquid-based
Juliana Crucello1, Danilo V Pierone2, Leandro W Hantao1
1Institute of Chemistry, University of Campinas, Campinas, SP 13083-970, Brazil.
Abstract:
In this article we describe a method using flow-modulated comprehensive two-dimensional gas chromatography with electron capture detector (FM-GC×GC-ECD) for the determination of polychlorinated biphenyls (PCBs) in complex transformer oils bypassing the need for sample preparation. A two-fold improvement in method development was attained. First, the solvation parameter model (SPM) was used to guide column selection. A highly cohesive ionic liquid-based phase (low l system constant), namely 1,12-di(tripropylphosphonium)dodecane bis(trifluoromethanesulfonyl)imide, was used in the primary stage leading to negligible retention of interfering aliphatic hydrocarbons, which are eluted in the first upper quadrant of the chromatogram. The resulting separation space was used to resolve the critical class of compounds, namely, PCBs and polyaromatic hydrocarbons. Second, a unique combination of column geometries and phase ratios enabled highly efficient reverse fill/flush flow modulation using very low pressure for auxiliary gas flow. The proof of concept method described herein exhibited linearities ranging from 0.990 to 0.994, limits of quantitation (LOQ) from 2.23 and 6.85 µg mL-1, precision below 5% relative standard deviation (RSD), and accuracy from 84.2% to 108.9% showcasing the potential of FM-GC×GC for routine analysis.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Gas Chromatography (GC) with Flame-Ionization Detection
Gas chromatography (GC) is used to separate and detect small molecular weight compounds in the gas phase. The sample is either a gas or a liquid that is vaporized in the injection port. Typically, the compounds analyzed are less than 1,000 Da, because it is difficult to vaporize larger compounds. GC is popular for environmental monitoring and industrial applications because it is very reliable and can be run nearly continuously....
10:42Pretreatment of Lignocellulosic Biomass with Low-cost Ionic Liquids
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection