Field drift correction of proton resonance frequency shift temperature mapping with multichannel fast alternating
Cyril J Ferrer1, Lambertus W Bartels1, Tijl A van der Velden1
1Imaging Division, University Medical Center Utrecht, Utrecht, Netherlands.
Purpose:
To demonstrate that proton resonance frequency shift MR thermometry (PRFS-MRT) acquisition with nonselective free induction decay (FID), combined with coil sensitivity profiles, allows spatially resolved B0 drift-corrected thermometry.
Methods:
Phantom experiments were performed at 1.5T and 3T. Acquisition of PRFS-MRT and FID were performed during MR-guided high-intensity focused ultrasound heating. The phase of the FIDs was used to estimate the change in angular frequency δωdrift per coil element. Two correction methods were investigated: (1) using the average δωdrift over all coil elements (0th-order) and (2) using coil sensitivity profiles for spatially resolved correction. Optical probes were used for independent temperature verification. In-vivo feasibility of the methods was evaluated in the leg of 1 healthy volunteer at 1.5T.
Results:
In 30 minutes, B0 drift led to an apparent temperature change of up to -18°C and -98°C at 1.5T and 3T, respectively. In the sonicated area, both corrections had a median error of 0.19°C at 1.5T and -0.54°C at 3T. At 1.5T, the measured median error with respect to the optical probe was -1.28°C with the 0th-order correction and improved to 0.43°C with the spatially resolved correction. In vivo, without correction the spatiotemporal median of the apparent temperature was at -4.3°C and interquartile range (IQR) of 9.31°C. The 0th-order correction had a median of 0.75°C and IQR of 0.96°C. The spatially resolved method had the lowest median at 0.33°C and IQR of 0.80°C.
Conclusion:
FID phase information from individual receive coil elements allows spatially resolved B0 drift correction in PRFS-based MRT.
More Related Videos
10:28Measuring Interactions of Globular and Filamentous Proteins by Nuclear Magnetic Resonance Spectroscopy NMR and Microscale Thermophoresis MST
Published on: November 2, 2018
07:01Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
Published on: June 9, 2016
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences