Impact of torsional micropulse on phacoemulsification efficiency and chatter
Sami W Kabbara1, Joshua Heczko2, Brian Ta2
1Banner-University of Arizona College of Medicine, Phoenix, AZ.
Objective:
Evaluate the effect of increasing ultrasound (US) power on chatter events and efficiency under both continuous and micropulse torsional US to reduce total cataract extraction times.
Design:
In vitro laboratory study.
Methods:
Porcine lens nuclei were incubated in formalin for 2 hours and then cut into 2-mm cubes. Phacoemulsification was performed using the Centurion Vision System and Infiniti OZil handpiece with the balanced tip. Both US modalities were studied at 60%, 80%, and 100% power. Micropulse rate was 83 pulses per second with 50% on time. Each combination comprised 20 runs. Efficiency was considered as the total time for a cube to be emulsified; chatter was the number of times the lens fragment bounced off the tip.
Results:
There was significant decrease in efficiency when power was increased from 60% to 100% (1.33-1.97 s; p < 0.001) under micropulse US and significant increase in chatter when power was further increased to 100% from 60% (0.15-0.94 s; p < 0.001). There was no significant efficiency change with increased power under continuous US. Comparing the phacoemulsification efficiency between continuous and micropulse US, we found no significant difference at 60% and 80% power; at 100% power, continuous was significantly more efficient than micropulse (1.48 and 1.97 s, respectively; p = 0.001).
Conclusions:
Increasing power above 60% decreased efficiency under torsional micropulse US. We believe that this was due to the chatter increase observed with increasing US power. Torsional continuous US was significantly more efficient than micropulse US at 100% power.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
08:12Author Spotlight: Exploring Light-Driven Chemical Reactions and Energy-Harnessing Devices in Photochemical Research
Published on: February 16, 2024
Related Concept Videos
Torsional Pendulum
As long as the rigid body's angular displacement is small, its oscillation can be modeled as a linear angular oscillation. The amplitude of the oscillation is an angle. The role of mass is played...
Residual Stresses in Circular Shafts
Stress Concentrations in Circular Shafts