Evaluation of additives on reversed-phase chromatography of monoclonal antibodies using a 1000 Å stationary phase
Related Experiment Video
Updated: Jan 6, 2026

Purification and Analytics of a Monoclonal Antibody from Chinese Hamster Ovary Cells Using an Automated Microbioreactor System
Published on: May 1, 2019
David V McCalley1, Davy Guillarme2
1Centre for Research in Biosciences, University of the West of England, Frenchay, Bristol BS16 1QY, UK.
Abstract:
A wide pore (1000 Å) diphenyl stationary phase was evaluated for the analysis of monoclonal antibodies (mAbs), comparing a conventional mobile phase of acetonitrile-water containing overall 0.1% trifluoracetic acid (TFA) with a similar mobile phase incorporating in addition 5% butanol. Alternatively, TFA was replaced by ammonium formate (AF) buffer (pH 3.0) and by methane sulfonic acid. Addition of 5% butanol to the mobile phase reduces the minimum temperature at which suitable UV analysis of the mAbs can be obtained from about 70 °C with TFA alone to about 60 °C thus potentially improving column lifetime and reducing the possibility of sample degradation. AF buffers produce satisfactory UV sensitivity at 70 °C and have the advantage of reducing signal suppression in mass spectrometry (MS). Some peak tailing was noted in comparison with TFA separations. Methane sulfonic acid at the same molar concentration as TFA produced the best chromatographic peaks, maintaining reasonable UV sensitivity down to 50 °C, also giving acceptable results even at only 3 mM concentration of the additive. The good results with this additive were attributed to its stronger acidity and consequent suppression of the ionisation of column silanols. Surprisingly, peak response (as measured by the size of the peaks) was rather poorly correlated with the peak capacity of the gradient analysis. A possible explanation is self-deactivation of active column sites by a portion of the sample.
More Related Videos
Related Concept Videos
Affinity Chromatography
Analyte Adsorption and Distribution
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...
Chromatographic Methods: Terminology
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...