Design and performance of closed cycle sample cooling stage for angle resolved photoemission spectroscopy capable of reaching temperatures below 2 K
Related Experiment Video
Updated: Jan 6, 2026

Angle-resolved Photoemission Spectroscopy At Ultra-low Temperatures
Published on: October 9, 2012
Benjamin Schrunk1, Lunan Huang1, Yun Wu1
1Division of Materials Science and Engineering, Ames Laboratory, Ames, Iowa 50011, USA.
Abstract:
We have designed, constructed, and tested a unique cold finger suitable for angle resolved photoemission spectroscopy. This design is based on in situ helium reliquification and utilizes pulse tube cryocooler. The pulse tube can be removed for baking without breaking Ultra High Vacuum (UHV). This design also allows the use of non-UHV heater that can be replaced without the need to vent the system. The cold finger has minimal vibration, operates over a temperature range of 1.7 K-400 K, and has no measurable residual magnetization. In continuous mode, it can maintain a sample temperature of 2.6 K, while in single shot mode (by pumping on liquid helium), it can reach temperatures down to 1.8 K for a period of several hours.