Optimization of an enzyme-coupling method by spectrophotometer for serum adenosine deaminase: As a candidate
Liqiao Han1, Xiaoting Huang1, Jianbing Wang1
1Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China.
Analytical Biochemistry
|October 5, 2019
Summary
This study developed a standardized reference measurement procedure for serum adenosine deaminase (ADA) to improve the accuracy of routine diagnostic tests. The optimized method ensures reliable ADA measurements for clinical applications.
Area of Science:
- Biochemistry
- Clinical Chemistry
- Enzyme Assays
Background:
- Adenosine deaminase (ADA) is crucial for adenosine metabolism.
- Existing ADA measurement kits show variable results, hindering accurate clinical interpretation.
- Standardization of ADA assays is needed for reliable diagnostic outcomes.
Purpose of the Study:
- To optimize and validate a reference measurement procedure (RMP) for serum ADA.
- To establish a standardized method for routine ADA testing.
- To improve the accuracy and comparability of ADA measurements across different laboratories.
Main Methods:
- Enzyme-linked assay using ADA, purine-nucleoside-phosphorylase, xanthine-oxidase, and peroxidase.
- Optimization of the assay using Response Surface Methodology.
- Validation of the RMP through inter-laboratory comparisons and assessment of precision, linearity, and interference.
Main Results:
- The optimized RMP demonstrated excellent intra-assay (0.44%) and inter-assay (≤1.01%) precision.
- High linearity was observed up to 351.9 U/L (R² = 0.9998) with minimal interference (<5%).
- Inter-laboratory reproducibility was high (R² ≥ 0.9975), and a reference interval of 11.7-38.5 U/L was established.
Conclusions:
- The developed candidate RMP for serum ADA offers a reliable method for standardizing routine clinical assays.
- Calibration of a routine system using the RMP significantly improved its accuracy (deviation reduced from -55.9% to -3.7%).
- This standardized procedure has the potential to enhance the diagnostic utility of ADA measurements in clinical settings.