Related Experiment Video
Updated: Jan 6, 2026

Diagnosis of Neoplasia in Barrett’s Esophagus using Vital-dye Enhanced Fluorescence Imaging
Published on: May 11, 2014
Endoscopic detection and differentiation of esophageal lesions using a deep neural network
Masayasu Ohmori1, Ryu Ishihara2, Kazuharu Aoyama3
1Department of Gastrointestinal Oncology, Osaka International Cancer Institute, Osaka, Japan; Department of Gastroenterology and Hepatology, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences, Okayama, Japan.
Background And Aims:
Diagnosing esophageal squamous cell carcinoma (SCC) depends on individual physician expertise and may be subject to interobserver variability. Therefore, we developed a computerized image-analysis system to detect and differentiate esophageal SCC.
Methods:
A total of 9591 nonmagnified endoscopy (non-ME) and 7844 ME images of pathologically confirmed superficial esophageal SCCs and 1692 non-ME and 3435 ME images from noncancerous lesions or normal esophagus were used as training image data. Validation was performed using 255 non-ME white-light images, 268 non-ME narrow-band images/blue-laser images, and 204 ME narrow-band images/blue-laser images from 135 patients. The same validation test data were diagnosed by 15 board-certified specialists (experienced endoscopists).
Results:
Regarding diagnosis by non-ME with narrow-band imaging/blue-laser imaging, the sensitivity, specificity, and accuracy were 100%, 63%, and 77%, respectively, for the artificial intelligence (AI) system and 92%, 69%, and 78%, respectively, for the experienced endoscopists. Regarding diagnosis by non-ME with white-light imaging, the sensitivity, specificity, and accuracy were 90%, 76%, and 81%, respectively, for the AI system and 87%, 67%, and 75%, respectively, for the experienced endoscopists. Regarding diagnosis by ME, the sensitivity, specificity, and accuracy were 98%, 56%, and 77%, respectively, for the AI system and 83%, 70%, and 76%, respectively, for the experienced endoscopists. There was no significant difference in the diagnostic performance between the AI system and the experienced endoscopists.
Conclusions:
Our AI system showed high sensitivity for detecting SCC by non-ME and high accuracy for differentiating SCC from noncancerous lesions by ME.