Related Experiment Video For Amperometric separation detection
Updated: Jan 6, 2026

Clinical Microfluidic Chip Platform for the Isolation of Versatile Circulating Tumor Cells
Published on: October 13, 2023
Separation detection of different circulating tumor cells in the blood using an electrochemical microfluidic channel
N G Gurudatt1, Saeromi Chung2, Jung-Min Kim2
1Department of Chemistry and Institute of BioPhysio Sensor Technology, Pusan National University, Busan, 46241, Republic of Korea; National Efficacy Evaluation Center for Metabolic Disease Therapeutics, Lee Gil Ya Cancer and Diabetes Institute, Gil Medical Center, Gachon University, Incheon, 21565, Republic of Korea.
Abstract:
Different circulating tumor cells (CTCs) in blood were separated and detected through the decoration of anti-cancer drug on the target cells, along with chemical modification of the microfluidic channel walls using a lipid attached covalently to the conducting polymer. The working principle of the electrochemical microfluidic device was evaluated with experimental parameters affecting on the separation, in terms of mass and surface charge of target species, fluid flow rate, AC amplitude, and AC frequency. The separated CTCs were selectively detected via the oxidation of daunomycin adsorbed specifically at the cells using an electrochemical sensor installed at the channel end. The fluorescence microscopic examination also confirmed the separation of CTCs in the channel. To evaluate the reliability of the method, blood samples from 37 cancer patients were tested. The device was able to separate the CTCs with 92.0 ± 0.5 % efficiency and 90.9% detection rate.