Orthopaedic metallic artefact reduction algorithm facilitates CT evaluation of the urinary tract after hip prosthesis
P N Kaposi1, T Youn1, A Tóth1
1Department of Radiology, Semmelweis University, Budapest, Hungary.
Aim:
To examine the improvement in the visualisation of bladder and ureteric pathologies next to a hip prosthesis with metallic artefact reduction for orthopaedic implants (O-MAR).
Materials And Methods:
Thirty-four patients who underwent pelvic computed tomography (CT) for non-prosthesis-related causes were enrolled retrospectively. Portal venous phase scans were reconstructed both with standard iterative reconstruction (ITR) and with O-MAR. The density of the ureters and the bladder was measured at both sides in the plane of the prosthesis. A semi-quantitative score was also used to assess visibility. The R (version 3.4.1) package was used for statistical analysis.
Results:
The average (μ) density of the 41 prosthesis side ureters was significantly lower on ITR images (μ=-94.76±150.48 [±SD] HU) than on O-MAR images (μ=-13.40±36.37 HU; p<0.0004). The difference between the ITR and O-MAR (μ=-138.62±182.64 versus -35.55±40.21 HU; p<0.0003) was also significant at the prosthesis side of the bladder. The visibility of the prosthesis side ureters was improved: 53.7% was obscured on ITR series compared to 4.9% on O-MAR. The visibility score was also better across all levels (p<0.001) with O-MAR. In four cases (13%), the O-MAR images significantly changed the diagnosis: in two cases ureteric stones, in one case each a bladder stone and a bladder tumour were discovered.
Conclusions:
O-MAR reconstruction of CT images significantly improves the visibility of the urinary tract adjacent to metallic hip implants. Thus, O-MAR is essential for detecting ureteric and bladder pathologies in patients with a hip prosthesis.