Related Experiment Video
Updated: Jan 5, 2026

Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Three-dimensional trueness analysis of ceramic crowns fabricated using a chairside computer-aided
Seen-Young Kang1, Jung-Hyun Park1, Ji-Hwan Kim1
1Department of Dental Laboratory Science and Engineering, College of Health Science, Korea University, 145, Anam-ro, Seoungbuk-gu, Seoul 136-712, Republic of Korea.
Purpose:
This study analyzed the trueness of polymer-infiltrated ceramic and glass ceramic crowns manufactured using the chairside computer-aided design/manufacturing (CAD/CAM) system.
Methods:
The master model designs crowns using a CAD program after acquiring a digital impression with an intraoral scanner. Vita Enamic (VE), Vita Suprinity (VS), and IPS e.max CAD (IPS) were used to manufacture 10 crowns each (total: 30 crowns), using the chairside CAD/CAM system (inLab MC XL). Trueness was evaluated by superimposing the CAD data on the scan data using a three-dimensional program. The Kruskal-Wallis H test, a nonparametric test, and the Mann-Whitney U test were performed by applying the significance level (0.05/3=0.016), which was adjusted by post-analysis Bonferroni testing.
Results:
There was a significant difference in the trueness between the samples (p<0.05). However, there was no statistically significant difference in the outer surface trueness between the samples (p>0.05).
Conclusions:
These findings show that the milling accuracy of VE is better than that of VS and IPS.