Characterizing Insertion Pressure Profiles During Cochlear Implantation: Simultaneous Fluoroscopy and Intracochlear
Joseph R Gonzalez1, Nathan D Cass1, Renee M Banakis Hartl1
1Department of Otolaryngology.
Summary
Cochlear implant electrode insertion generates loud pressure transients in the inner ear. Identifying the source of these pressure transients is key to developing quieter insertion techniques for better hearing preservation in patients undergoing cochlear implantation.
Area of Science:
- Otoacoustic Emissions
- Cochlear Implant Technology
- Auditory Neuroscience
Background:
- Combined electrical-acoustical stimulation (EAS) is increasingly used in cochlear implantation for patients with residual hearing.
- Hearing preservation is crucial for cochlear implant outcomes, but some patients experience hearing loss.
- High-amplitude pressure transients, akin to noise exposure, occur during electrode insertion, but their source is unknown.
Purpose of the Study:
- To investigate the source of high-amplitude pressure transients during cochlear implant electrode insertion.
- To analyze intracochlear pressure changes associated with different electrode insertion techniques.
- To correlate pressure transient generation with electrode position within the cochlea.
Main Methods:
- Cadaveric human heads were used with extended facial recess preparation.
- Fiber-optic pressure sensors were placed in the scala vestibuli and scala tympani.
- Two cochlear implant electrode styles (straight and perimodiolar) were inserted while measuring intracochlear pressures and using video fluoroscopy.
Main Results:
- Cochlear implant electrode insertions generated pressure transients up to 160-170 dB pSPL for both electrode styles.
- The specific location within the cochlea where transients occurred varied, especially with contact to the medial or lateral walls.
Conclusions:
- This study elucidates the pressure profiles generated during cochlear implant electrode insertion.
- Findings can inform improved cochlear implant electrode designs.
- Development of "silent-insertion" techniques may enhance hearing preservation.