The measurement of blood pressure by the linear method compared to the deflation methods differently modifies the
Norihisa Kitamoto1,2, Takashi Hitosugi1, Yoshifumi Kawakubo1
1Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University, Fukuoka.
Background:
Noninvasive blood-pressure measurement device and pulse oximeter are important for patient monitoring. When these are placed on the same side, cuff inflation sometimes causes measurement failure by pulse oximeter.
Objective:
The present study aimed to compare the pulse oximeter alarm frequency and pulse-wave disappearance duration between noninvasive blood-pressure measurement using the deflation method and that using the linear inflation method.
Methods:
The study included 10 healthy subjects. The cuff for automatic sphygmomanometer was wrapped on one side of the upper arm and for pulse oximeter was attached to the thumb of the same side of upper limbs.
Results:
The alarm frequency was 0 and 26% using the linear inflation and the deflation methods, respectively. Additionally, the pulse-wave disappearance duration was significantly longer using the deflation method than that using the linear inflation method (10.0 ± 1.5 vs 1.7 ± 0.8 s). With the linear inflation method, this duration was or less 3 s. In the deflation method, an excess pressure of 40 mmHg was used, which caused the alarm to turn on. Additionally, the heart rate was found to influence the alarm occurrence during measurement using the deflation method.
Conclusion:
Heart rate may influence alarm occurrence during blood-pressure measurement using the step deflation method. Using the linear inflation method, the risks of alarm occurrence and measurement failure are low, even when the pulse oximeter and blood-pressure measurement cuffs are installed on the same side, suggesting that this method is suitable for clinical use.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
11:26Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
Published on: December 10, 2014
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...