Method Verification Shows a Negative Bias between 2 Procalcitonin Methods at Medical Decision Concentrations
Valentinas Gruzdys1, Kenneth Cahoon2, Lauren Pearson3
1Department of Pathology and Laboratory Medicine, Loyola University Medical Center, Maywood, IL; valentinas.gruzdys@lumc.edu.
Background:
Procalcitonin (PCT) concentration increases as a result of systemic inflammation owing to bacterial infection. Many PCT algorithms and medical decision concentrations (MDCs) have been clinically validated using the B·R·A·H·M·S PCT™ sensitive Kryptor™ assay. Alternative PCT assays have recently been approved by the Food and Drug Administration for clinical use in the US and require method verification before clinical implementation.
Methods:
Precision, sensitivity, linearity, reportable range, and reference intervals were verified for the Architect B·R·A·H·M·S PCT assay. Accuracy of the Architect B·R·A·H·M·S PCT assay was evaluated by comparison with the B·R·A·H·M·S PCT sensitive Kryptor assay.
Results:
The Architect B·R·A·H·M·S PCT assay was found to be precise (CV, ≤4.6%), sensitive (limit of blank, 0.001 ng/mL; limit of quantitation, ≤0.01 ng/mL), and linear according to the manufacturer's claims. The analytical measurement range (0.20-100.00 ng/mL) and the reference interval (≤0.07 ng/mL) were also verified. Patient result comparisons indicated high agreement at 0.10 ng/mL and 0.25 ng/mL and reduced positive agreement at 0.50 ng/mL and 2.00 ng/mL MDCs owing to negative bias compared with the B·R·A·H·M·S PCT sensitive Kryptor assay.
Conclusions:
The Architect B·R·A·H·M·S PCT assay meets most performance specifications claimed by the manufacturer; however, negative bias at 0.50 ng/mL and 2.00 ng/mL PCT concentrations is evident.
More Related Videos
07:52Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
16:07Determination of the Relative Potency of an Anti-TNF Monoclonal Antibody mAb by Neutralizing TNF Using an In Vitro Bioanalytical Method
Published on: September 16, 2017
Related Concept Videos
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...
Therapeutic Drug Monitoring: Drug Analysis Methods
Data Validation
Key parameters for method validation include: