[Determination of chlorate and perchlorate in milk power by high-performance liquid chromatography-tandem mass
Xiaoqing Zhou1, Xiaoli Lü1, Jianchun Wan1
1Jiang Xi Institute for Food Control, Nanchang 330001, China.
Abstract:
An analytical method based on high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) was developed for detecting chlorate and perchlorate residues in milk power. Chlorate and perchlorate in milk power were extracted using a 0.1% (v/v) formic acid-acetonitrile solution. The extract was centrifuged at 10000 r/min for 10 min, and the supernatant was cleaned up on a PRiME HLB column. Separation of chlorate and perchlorate was performed on an ion-exchange column (Thermo Scientific Acclaim TRINITY P1, 50 mm×2.1 mm, 3 μm) by gradient elution using acetonitrile and 20 mmol/L ammonium acetate solution as the mobile phase. The analytes were identified by MS/MS. Quantification was achieved using internal standards. Chlorate and perchlorate demonstrated good linearity in the ranges of 2.0-40.0 and 1.0-20.0 μg/L, respectively, with correlation coefficients (r2) greater than 0.999. The limits of quantification (LOQs) of chlorate and perchlorate were found to be 15.0 and 7.5 μg/kg, respectively. The recoveries of chlorate and perchlorate ranged from 89.24% to 107.85% at the three spike levels of 30.0, 60.0, and 120.0, and 15.0, 30.0, and 60.0 μg/kg, respectively, with relative standard deviations (RSDs) ranging from 3.15% to 10.42% (n=6). This method is convenient, rapid, accurate, and efficient, thus demonstrating its suitability for use in the determination of chlorate and perchlorate in milk power.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Qualitative Analysis
For instance, group IV...
Chromatographic Methods: Terminology