Related Experiment Video
Updated: Jan 5, 2026

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Glottic visibility for laryngeal surgery: Tritube vs. microlaryngeal tube: A randomised controlled trial
Johannes Schmidt1, Franziska Günther, Jonas Weber
1From the Department of Anaesthesiology and Critical Care (JS, FG, JW, VK, CW, SB, SW, SS) and Department of Otorhinolaryngology (JP, CB), Medical Centre - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Background:
Good visibility is essential for successful laryngeal surgery. A Tritube with outer diameter 4.4 mm, combined with flow-controlled ventilation (FCV), enables ventilation by active expiration with a sealed trachea and may improve laryngeal visibility.
Objectives:
We hypothesised that a Tritube with FCV would provide better laryngeal visibility and surgical conditions for laryngeal surgery than a conventional microlaryngeal tube (MLT) with volume-controlled ventilation (VCV).
Design:
Randomised, controlled trial.
Setting:
University Medical Centre.
Patients:
A total of 55 consecutive patients (>18 years) undergoing elective laryngeal surgery were assessed for participation, providing 40 evaluable data sets with 20 per group.
Interventions:
Random allocation to intubation with Tritube and ventilation with FCV (Tritube-FCV group) or intubation with MLT 6.0 and ventilation with VCV (MLT-VCV) as control. Tidal volumes of 7 ml kg predicted body weight, and positive end-expiratory pressure of 7 cmH2O were standardised between groups.
Main Outcome Measures:
Primary endpoint was the tube-related concealment of laryngeal structures, measured on videolaryngoscopic photographs by appropriate software. Secondary endpoints were surgical conditions (categorical four-point rating scale), respiratory variables and change of end-expiratory lung volume from atmospheric airway pressure to ventilation with positive end-expiratory pressure. Data are presented as median [IQR].
Results:
There was less concealment of laryngeal structures with the Tritube than with the MLT; 7 [6 to 9] vs. 22 [18 to 27] %, (P < 0.001). Surgical conditions were rated comparably (P = 0.06). A subgroup of residents in training perceived surgical conditions to be better with the Tritube compared with the MLT (P = 0.006). Respiratory system compliance with the Tritube was higher at 61 [52 to 71] vs. 46 [41 to 51] ml cmH2O (P < 0.001), plateau pressure was lower at 14 [13 to 15] vs. 17 [16 to 18] cmH2O (P < 0.001), and change of end-expiratory lung volume was higher at 681 [463 to 849] vs. 414 [194 to 604] ml, (P = 0.023) for Tritube-FCV compared with MLT-VCV.
Conclusion:
During laryngeal surgery a Tritube improves visibility of the surgical site but not surgical conditions when compared with a MLT 6.0. FCV improves lung aeration and respiratory system compliance compared with VCV.
Trial Registry Number:
DRKS00013097.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...