Activity determination of 231Pa by means of liquid scintillation counting
Haoran Liu1, Qianqian Zhou2, Fuyou Fan2
1National Institute of Metrology (NIM), Beijing, 100029, China; China Institute of Atomic Energy (CIAE), Beijing, 102413, China.
Abstract:
Protactinium-231 decays through a chain of daughter nuclides to stable 207Pb, resulting in a number of alpha and beta transitions. A method based on liquid scintillation (LS) counting was developed to determine the activity concentration of 231Pa. For α emissions, the corresponding LS counting efficiency was assumed to be unity. A potential loss of the noble gas 219Rn was evaluated by an additional experiment. In the case of the short-lived 215Po, a dead-time-dependent correction factor was taken into account in the calculation. For beta emitters 227Ac, 223Fr, 211Pb and 207Tl, the counting efficiencies were determined by the CIEMAT/NIST efficiency tracing (CNET) method with 3H standard as a tracer. The impurities were checked and found that the possible impurity was 227Ac, with an initial activity ratio A(227Ac)/A (231Pa) between 0.5% and 0.8%. Finally, the activity concentration of the 231Pa solution was determined with a combined relative standard uncertainty of 0.39%.
More Related Videos
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview