A noise correction of the γ-index method for Monte Carlo dose distribution comparison
Marie Cohilis1, Edmond Sterpin1,2, John A Lee1
1Center of Molecular Imaging, Radiotherapy and Oncology (MIRO), Institut de Recherche Expérimentale et Clinique (IREC), Université catholique de Louvain, 1200, Brussels, Belgium.
Purpose:
Due to the increasing complexity of IMRT/IMPT treatments, quality assurance (QA) is essential to verify the quality of the dose distribution actually delivered. In this context, Monte Carlo (MC) simulations are more and more often used to verify the accuracy of the treatment planning system (TPS). The most common method of dose comparison is the γ-test, which combines dose difference and distance-to-agreement (DTA) criteria. However, this method is known to be dependent on the noise level in dose distributions. We propose here a method to correct the bias of the γ passing rate (GPR) induced by MC noise.
Methods:
The GPR amplitude was studied as a function of the MC noise level. A model of this noise effect was mathematically derived. This model was then used to predict the time-consuming low-noise GPR by fitting multiple fast MC dose calculations. MC dose maps with a noise level between 2% and 20% were computed, and the GPR was predicted at a noise level of 0.3%. Due to the asymmetry of the γ-test, two different cases were considered: the MC dose was first set as reference dose, then as evaluated dose in the γ-test. Our method was applied on six proton therapy plans including analytical doses from the TPS or patient-specific QA measurements.
Results:
An average absolute error of 4.31% was observed on the GPR computed for MC doses with 2% statistical noise. Our method was able to improve the accuracy of the gamma passing rate by up to 13%. The method was found especially efficient to correct the noise bias when the DTA criterion is low.
Conclusions:
We propose a method to enhance the γ-evaluation of a treatment plan when there is noise in one of the compared distributions. The method allows, in a tractable time, to detect the cases for which a correction is necessary and can improve the accuracy of the resulting passing rates.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Maxwell-Boltzmann Distribution: Problem Solving
This distribution function f(v) is defined by saying that the expected number N (v1,v2) of particles with speeds between v1 and v2 is given by
Dose Size and Dosing Frequency: Determination Methods
Biological Effects of Radiation
NMR Spectrometers: Resolution and Error Correction
Propagation of Uncertainty from Systematic Error
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force...