Segmentation and Linear Measurement for Body Composition Analysis using Slice-O-Matic and Horos
Published on: March 21, 2021
Reproducibility of semiautomated body composition segmentation of abdominal computed tomography: a multiobserver
Lisa Jannicke Kjønigsen1, Magnus Harneshaug2,3, Ann-Monica Fløtten1
1Division of Radiology and Nuclear Medicine, Oslo University Hospital, Oslo, Norway.
Background:
Segmentation of computed tomography (CT) images provides quantitative data on body tissue composition, which may greatly impact the development and progression of diseases such as type 2 diabetes mellitus and cancer. We aimed to evaluate the inter- and intraobserver variation of semiautomated segmentation, to assess whether multiple observers may interchangeably perform this task.
Methods:
Anonymised, unenhanced, single mid-abdominal CT images were acquired from 132 subjects from two previous studies. Semiautomated segmentation was performed using a proprietary software package. Abdominal muscle compartment (AMC), inter- and intramuscular adipose tissue (IMAT), visceral adipose tissue (VAT) and subcutaneous adipose tissue (SAT) were identified according to pre-established attenuation ranges. The segmentation was performed by four observers: an oncology resident with extensive training and three radiographers with a 2-week training programme. To assess interobserver variation, segmentation of each CT image was performed individually by two or more observers. To assess intraobserver variation, three of the observers did repeated segmentations of the images. The distribution of variation between subjects, observers and random noise was estimated by a mixed effects model. Inter- and intraobserver correlation was assessed by intraclass correlation coefficient (ICC).
Results:
For all four tissue compartments, the observer variations were far lower than random noise by factors ranging from 1.6 to 3.6 and those between subjects by factors ranging from 7.3 to 186.1. All interobserver ICC was ≥ 0.938, and all intraobserver ICC was ≥ 0.996.
Conclusions:
Body composition segmentation showed a very low level of operator dependability. Multiple observers may interchangeably perform this task with highly reproducible results.