An automated approach for the optimised estimation of breast density with Dixon methods
Rosie Goodburn1, Evanthia Kousi1, Alison Macdonald2
1CRUK Cancer Imaging Centre, The Institute of Cancer Research and Royal Marsden Foundation Trust, London, United Kingdom.
Objective:
To present and evaluate an automated method to correct scaling between Dixon water/fat images used in breast density (BD) assessments.
Methods:
Dixon images were acquired in 14 subjects with different T1 weightings (flip angles, FA, 4°/16°). Our method corrects intensity differences between water () and fat () images via the application of a uniform scaling factor (SF), determined subject-by-subject. Based on the postulation that optimal SFs yield relatively featureless summed fat/scaled-water () images, each SF was chosen as that which generated the lowest 95th-percentile in the absolute spatial-gradient image-volume of . Water-fraction maps were calculated for data acquired with low/high FAs, and BD (%) was the total percentage water within each breast volume.
Results:
Corrected/uncorrected BD ranged from, respectively, 10.9-71.8%/8.9-66.7% for low-FA data to 8.1-74.3%/5.6-54.3% for high-FA data. Corrected metrics had an average absolute increase in BD of 6.4% for low-FA data and 18.4% for high-FA data. BD values estimated from low- and high-FA data were closer following SF-correction.
Conclusion:
Our results demonstrate need for scaling in such BD assessments, where our method brought high-FA and low-FA data into closer agreement.
Advances In Knowledge:
We demonstrated a feasible method to address a main source of inaccuracy in Dixon-based BD measurements.