Related Experiment Video
Updated: Jan 4, 2026

Use of the Scissor-Type Knife During the Peroral Endoscopy Myotomy Procedure for the Treatment of Achalasia
Published on: March 3, 2023
Novel single-use duodenoscope compared with 3 models of reusable duodenoscopes for ERCP: a randomized bench-model
Andrew S Ross1, Marco J Bruno2, Richard A Kozarek1
1Department of Gastroenterology, Digestive Disease Institute, Virginia Mason Medical Center, Seattle, Washington, USA.
Background And Aims:
Multidrug-resistant infectious outbreaks associated with duodenoscope reuse have been documented internationally. A single-use endoscope could eliminate exogenous patient-to-patient infection associated with ERCP.
Methods:
We conducted a comparative bench simulation study of a new single-use and 3 models of reusable duodenoscopes on a synthetic anatomic bench model. Four ERCP tasks were performed: guidewire locking (single-use and 1 reusable duodenoscope only), plastic stent placement and removal, metal stent placement and removal, and basket sweeping. The study schedule included block randomization by 4 duodenoscopes, 4 tasks, and 2 anatomic model ERCP stations. Ability to complete tasks, task completion times, and subjective ratings of overall performance, navigation/pushability, tip control, and image quality on a scale of 1 (worst) to 10 (best) were compared among duodenoscopes.
Results:
All 4 ERCP tasks (total 14 subtasks) were completed by 6 expert endoscopists using all 4 duodenoscopes, with similar task completion times (median, 1.5-8.0 minutes per task) and overall performance ratings by task (median, 8.0-10.0). Navigation/pushability ratings were lower for the single-use duodenoscope than for the 3 reusable duodenoscopes (median, 8.0, 10.0, 9.0, and 9.0, respectively; P < .01). Tip control ratings were similar among all the duodenoscopes (median, 9.0-10.0; P = .77). Image quality ratings were lower for 1 reusable duodenoscope compared with the single-use and other 2 reusable duodenoscopes (median, 8.0, 9.0, 9.0, and 9.0, respectively; P < .01).
Conclusions:
A new single-use duodenoscope was used to simulate 4 ERCP tasks in an anatomic model, with performance ratings and completion times comparable with 3 models of reusable duodenoscopes.
Related Concept Videos
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...