Image-stitching artefacts and distortion in CCD-based cephalograms and their association with sensor type and head
Olesya Svystun1, Ann Wenzel1, Lars Schropp1
1Department of Dentistry and Oral Health, Oral Radiology, Aarhus University, Vennelyst Boulevard 9, 8000, Aarhus, Denmark.
Objectives:
To assess presence and severity of image-stitching artefacts and distortion in lateral cephalograms acquired by CCD-based sensors and their association with movement.
Methods:
A human skull was mounted on a robot simulating five head movement types (anteroposterior translation/lifting/nodding/lateral rotation/tremor), at three distances (0.75/1.5/3 mm), based on two patterns (skull returning/not returning to the initial position, except for tremor). Three cephalometric units, two ProMax-2D (Planmeca Oy, Finland), one with Dimax-3 (D-3) and one with Dimax-4 (D-4) sensor, and one Orthophos-SL (ORT, Dentsply-Sirona, Germany), acquired cephalograms during the predetermined movements, in duplicate (54 with movement and 28 controls with no movement per unit). One observer assessed the presence of an image-stitching line (none/thin/thin with vertical stripes or thick), misalignment between the anatomical structure display (none/<1/1-3/>3 mm), and distortion in each image quadrant (present/absent), in duplicate. Severe image-stitching artefacts were defined for images scored with a thin line with vertical stripes or thick line and/or misalignment between anatomical structure display ≥1 mm. Severe distortion was defined for images scored with distortion in both anterior quadrants of the skull. κ-statistics provided intraobserver agreement.
Results:
Intraobserver reproducibility was >0.8 (all assessed parameters). Severe image-stitching artefacts were scored in 70.4 and 18.5% of D-3 and D-4 movement images, respectively. Severe distortion was scored in 64.8% of D-3, 5.6% of D-4 and 37% of ORT movement images. Neither severe image-stitching artefacts nor severe distortion were observed in control images.
Conclusion:
Sensor type, movement type, distance and pattern affected presence and severity of image-stitching artefacts and distortion in CCD-based cephalograms.