Local impedance guides catheter ablation in patients with ventricular tachycardia
Paula Münkler1,2, Melanie A Gunawardene1,3, Christiane Jungen1,2
1Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
Aims:
Catheter contact and local tissue characteristics are relevant information for successful radiofrequency current (RFC)-ablation. Local impedance (LI) has been shown to reflect tissue characteristics and lesion formation during RFC-ablation. Using a novel ablation catheter incorporating three mini-electrodes, we investigated LI in relation to generator impedance (GI) in patients with ventricular tachycardia (VT) and its applicability as an indicator of effective RFC-ablation.
Methods And Results:
Baseline impedance, Δimpedance during ablation and drop rate (Δimpedance/time) were analyzed for 625 RFC-applications in 28 patients with recurrent VT undergoing RFC-ablation. LI was lower in scarred (87.0 Ω [79.0-95.0]) compared to healthy myocardium (97.5 Ω ([82.75-111.50]; P = .03) while GI did not differ between scarred and healthy myocardium. ΔLI was higher (18 Ω [9.4-26.0]) for VT-terminating as compared to non-terminating RFC-ablation (ΔLI 13 Ω [8.85-18.0]; P = .03), but did not differ for ΔGI between terminating vs nonterminating RFC-ablation. Correspondingly, LI drop rate was higher for RFC-ablation terminating the VT compared with RFC-ablation not terminating the VT (0.63 Ω/s [0.52-0.76] vs 0.32 Ω [0.20-0.58]; P = .008) while there was no difference for GI drop rate. ΔLI was higher in patients with nonischemic cardiomyopathy vs patients with ischemic cardiomyopathy (16 Ω [11.0-20.0] vs 11.0 Ω [7.85-17.00]; P = .003).
Conclusion:
Our findings suggest that LI is a sensitive parameter to guide RFC-ablation in patients with VT. LI indicates differences in tissue characteristics and generally is higher in patients with nonischemic cardiomyopathy. Hence, the etiology of the underlying cardiomyopathy needs to be considered when adopting LI for monitoring catheter ablation of VT.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiac Catheterization IV: Nursing Management
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiac Catheterization II: Right Heart Catheterization