Biomechanical comparison of bone-screw-fasteners versus traditional locked screws in plating female geriatric bone
Malcolm R DeBaun1, Steven T Swinford1, Michael J Chen1
1Department of Orthopaedics, School of Medicine, Stanford University, 450 Broadway St., MC 6342, Redwood City, CA 94063, United States.
Objectives:
To biomechanically compare plated constructs using nonlocking bone-screw-fasteners with interlocking threads versus locking screws with traditional buttress threads in geriatric female bone.
Methods:
Eleven matched pairs of proximal and distal segments of geriatric female cadaveric tibias were used to create a diaphyseal fracture model. Nonlocking bone-screw-fasteners or locking buttress threaded screws were applied to a locking compression plate on the anterolateral aspect of the tibia placed in bridge mode. Specimens were subjected to incrementally increasing cyclic axial load combined with constant cyclic torsion. Total cycles to failure served as a primary outcome measure, with failure defined as 2 mm of displacement or 10 degrees of rotation. Secondary outcome measures included initial stiffness in compression and torsion determined from preconditioning testing and overall rigidity as determined by maximum peak-to-peak axial and rotational motion at 500 cycle intervals during cyclic testing. Group comparisons were made using paired Student's t-tests. Significance was set at p < 0.05.
Results:
Bone-screw-fastener constructs failed at an average of 40,636 ± 22,151 cycles and locking screw constructs failed at an average of 37,773 ± 8433 cycles, without difference between groups (p = =0.610). Total cycles to failure was higher in the bone-screw-fasteners group for 7 tibiae out of the eleven matched pairs tested. During static and cyclic testing, bone-screw-fastener constructs demonstrated increased initial torsional stiffness (7.6%) and less peak-to-peak displacement and rotation throughout the testing cycle(p < 0.05).
Conclusions:
In female geriatric bone, constructs fixed with bone-screw-fasteners incorporate multiplanar interlocking thread geometry and performed similarly to traditional locked plating. These novel devices may combine the benefits of both nonlocking and locking screws when plating geriatric bone.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Wood Fasteners
Nails are the most common fasteners, consisting of sharp-pointed metal pins that are driven into wood using a hammer or a mechanical nail gun. They come in a variety of diameters, heads, and lengths. Nailing techniques include...