Cytological examination of cerebrospinal fluid: Sysmex UF-1000i versus optical microscopy
A Maleb1, O Bouayadi1, J El Malki1
1Laboratory of Microbiology, Mohammed VI University Hospital, Faculty of Medicine and Pharmacy (University Mohammed the First), Oujda, Morocco.
Abstract:
We evaluated the body fluid module on Sysmex UF-1000i (UF-1000i-BF) for analysis of white blood cell (WBC) and red blood cell (RBC) in cerebrospinal fluid. We collected 93 cerebrospinal fluid samples and compared the results of the UF-1000i-BF mode with the Fast-Read 102 disposable counting cell. Results shows a good correlation between the UF-1000i and the microscopic examination. The concordance percentage is 99.06% for white blood cells and 85.18% for red blood cells. The UF-1000i-BF mode offers rapid and reliable total WBC and RBC counts for initial screening of cerebrospinal fluid, and can improve the workflow in a routine laboratory.