Characterisation of a β detector on positron emitters for medical applications
F Collamati1, R Moretti2, L Alunni-Solestizi3
1Istituto Nazionale di Fisica Nucleare, Sezione di Roma, Roma, Italy.
Purpose:
Radio Guided Surgery (RGS) is a technique that helps the surgeon to achieve an as complete as possible tumor resection, thanks to the intraoperative detection of particles emitted by a radio tracer that bounds to tumoral cells. In the last years, a novel approach to this technique has been proposed that, exploiting β- emitting radio tracers, overtakes some limitations of established γ-RGS. In this context, a first prototype of an intraoperative β particle detector, based on a high light yield and low density organic scintillator, has been developed and characterised on pure β- emitters, like 90Y. The demonstrated very high efficiency to β- particles, together with the remarkable transparency to photons, suggested the possibility to use this detector also with β+ emitting sources, that have plenty of applications in nuclear medicine. In this paper, we present upgrades and optimisations performed to the detector to reveal such particles.
Methods:
Laboratory measurement have been performed on liquid Ga68 source, and were used to validate and tune a Monte Carlo simulation.
Results:
The upgraded detector has an ~80% efficiency to electrons above ~110keV, reaching a plateau value of ~95%. At the same time, the probe is substantially transparent to photons below ~200keV, reaching a plateau value of ~3%.
Conclusions:
The new prototype seems to have promising characteristics to perform RGS also with β+ emitting isotopes.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
09:08Enhancing Efficiency and Radiolabeling Yields of Carbon-11 Radioligands for Clinical Research Using the Loop Method
Published on: December 20, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Isotopes and Radioisotopes
An isotope containing...
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...