Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
A comparison of finger and forehead pulse oximeters in heart failure patients during maximal exercise
Kevin L Kelly1, Alex R Carlson2, Thomas G Allison2
1Mayo Clinic Department of Clinical and Translational Science and Activities, 200 1st Street SW, Rochester, MN 55905, United States; Mayo Clinic Department of Cardiovascular Diseases, 200 1st Street SW, Rochester, MN 55905, United States.
Background:
Pulse oximeters, clinically used to measure oxygen saturation (SpO2), rely on adequate perfusion of the tissues over which they are placed. Heart failure (HF) patients can have impaired peripheral perfusion which may compromise the accuracy of a peripherally placed pulse oximeter. This decrease in peripheral perfusion may be especially apparent during exercise. The objective of this study was to determine if pulse oximeter accuracy is dependent on location in heart failure patients during peak exercise.
Methods:
20 participants with HF (7F, age 64.±11 yr) and 9 participants with coronary artery disease as controls (CAD: 3F, age 66±5 yr) performed a maximal exertion treadmill exercise stress test while wearing both finger and forehead pulse oximeters.
Results:
At peak exercise, the two pulse oximeters measurements of SpO2 differed from each other by 3.8 ± 3.3% in the HF group (p<0.01) and 2.0 ± 1.4% in the CAD group (p = 0.065). The difference between the pulse rate from the pulse oximeters and the heart rate from the 12-lead ECG in the HF group was 12±20 BPM (p<0.01) for the finger pulse oximeter, and 2 ± 3 BPM (p = 0.162) for the forehead pulse oximeter.
Conclusions:
Forehead pulse oximeters may be more reliable compared to finger pulse oximeters in obtaining SpO2 measurements in HF patients during a treadmill maximal exercise test.
More Related Videos
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
09:04Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: