Discrepancies in Endothelial Cell Density Values of Human Donor Corneas Resulting From Comparison Between Specular
Yue Shi1, Jianyan Huang1,2, Patrick Conrick3
1Doheny Image Reading Center, Doheny Eye Institute, Los Angeles, CA.
Purpose:
The methods for specular microscopy evaluation across eye banks differ, which may result in variability in endothelial cell density (ECD) values that influence the surgeon's decision about donor tissue. A comparison of instruments and analysis methods is conducted in this study.
Methods:
Specular images were captured from 97 donor corneas using both HAI and Konan specular microscopes. A single best quality image of each cornea from each instrument was graded using the respective inherent software and analysis method (HAI: variable frame method; Konan: center method). All raw specular images were standardized for dimensions and regraded in the CellChek system in a blinded fashion. The grading variances and paired t test were performed between instruments in both inherent and standardized analyses. Correlation and Bland-Altman analyses between instruments were also performed.
Results:
Using the software inherent within HAI and Konan, the mean ECD readings for the 97 corneas were 2764 ± 583 and 2605 ± 517 cell/mm (P = 0.045), respectively, with a variance of 8.05% (range 0.26%-27.2%). HAI resulted in a higher ECD value in 79 corneas (81.4%). In CellChek software analysis, the mean ECD readings did not differ (2609 ± 514 and 2496 ± 507 cells/mm, respectively, P = 0.127), with a variance of 5.6% (range 0.24%-19.8%).
Conclusions:
There is a slight statistically significant mean difference between the ECD values obtained from the 2 specular microscopes, which is negated by standardization to a single analysis method. Eye banks and surgeons should use caution in making decisions based only on very small differences in ECD between otherwise equivalent corneal donor tissues.