Related Experiment Video
Updated: Jan 3, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Performance comparison of prediction filters for respiratory motion tracking in radiotherapy
Alexander Jöhl1,2, Stefanie Ehrbar2,3, Matthias Guckenberger2,3
1Product Development Group Zurich, Department of Mechanical and Process Engineering, ETH Zurich, Zurich, Switzerland.
Purpose:
In precision radiotherapy, the intrafractional motion causes substantial uncertainty. Traditionally, the target volume is expanded to cover the tumor in all positions. Alternative approaches are gating and adaptive tracking, which require a time delay as small as possible between the actual tumor motion and the reaction to effectively compensate the motion. Current treatment machines often exhibit large time delays. Prediction filters offer a promising means to mitigate these time delays by predicting the future respiratory motion.
Methods:
A total of 18 prediction filters were implemented and their hyperparameters optimized for various time delays and noise levels. A set of 93 traces were standardized to a sampling frequency of 25 Hz and smoothed using the Fourier transform with a 3 Hz cutoff frequency. The hyperparameter optimization was carried out with ten traces, and the optimal hyperparameters were evaluated on the remaining 83 traces.
Results:
For smooth traces, the wavelet least mean squares prediction filter and the linear filter reached normalized root mean square errors of below 0.05 for time delays of 160 and 480 ms, respectively. For noisy signals, the performance of the prediction filters deteriorated and led to similar results.
Conclusions:
Linear methods for prediction filters are sufficient for respiratory motion signals. Reducing the measurement noise generally improves the performance of the prediction filters investigated in this study, even during breathing irregularities.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018