Related Experiment Video
Updated: Jan 3, 2026

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Monitoring core temperature of firefighters to validate a wearable non-invasive core thermometer in different types
C C Roossien1, R Heus2, M F Reneman1
1University of Groningen, University Medical Center Groningen, Department of Rehabilitation Medicine, Hanzeplein 1, 9713 GZ, Groningen, the Netherlands.
Abstract:
This study aims (1) to test the validity of a new non-invasive core thermometer, Cosinuss°, in rest and (2) during firefighting simulation tasks, against invasive temperature pill and inner-ear temperature and (3) to compare the change in core temperature of firefighters when working in two types of protective clothing (traditional turnout gear versus new concept). 11 active firefighters performed twice a selection of tasks during their periodic preventive medical examination and a fire-extinguishing task. Without correction no correlation between the Cosinuss° and thermometer pill (ICC≤0.09, p ≥ 0.154, LoA≥1.37) and a moderate correlation between Cosinuss° and inner-ear infrared (ICC = 0.40, p = 0.044, LoA±1.20) was observed. With individual correction both correlations were excellent (ICC≥0.84, p = 0.000, LoA≤0.30). However, during and after working all correlations were poor and non-significant (ICC≤0.38, p ≥ 0.091, LoA≥1.71). During firefighting tasks, the Cosinuss° is invalid for measuring the core temperature. No differences in heat development in the two types of protective clothing was proven.
More Related Videos
05:32A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
Published on: November 24, 2016
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...