Environmental radiofrequency electromagnetic field levels in a department of pediatrics
Dimitri Besset1, Brahim Selmaoui1, Pierre Tourneux2
1PériTox-INERIS Laboratory UMR_I 01, Jules Verne University of Picardy, Amiens, France.
Insights
Radiofrequency electromagnetic field (RF-EMF) levels were measured in pediatric units, finding them very low overall. Exposure was higher during the day and in the neonatal critical care unit (NCCU), likely from wireless devices used by staff and parents.
Area of Science:
- Environmental Health
- Pediatrics
- Biophysics
Background:
- Preterm neonates are vulnerable to environmental factors.
- Increased use of wireless devices raises concerns about radiofrequency electromagnetic field (RF-EMF) exposure in neonatal intensive care units (NICUs) and neonatal critical care units (NCCUs).
- Previous data on RF-EMF levels in pediatric units were lacking.
Purpose of the Study:
- To quantify RF-EMF levels in a tertiary pediatric department, including NCCU and NICU.
- To assess RF-EMF emissions from medical devices used in neonatal care.
Main Methods:
- RF-EMF levels were measured using triaxle antenna dosimeters at 144 points across NCCU and NICU over 24 hours.
- Spot emissions from medical devices were measured in a Faraday chamber.
- Measurements were compared between NCCU and NICU, and between day and night.
Main Results:
- Overall RF-EMF levels in NCCU and NICU were very low.
- RF-EMF levels were significantly higher in the NCCU than in the NICU.
- Exposure levels were significantly higher during the day than at night.
- Medical devices did not emit detectable RF, suggesting exposure primarily from wireless devices used by people.
Conclusions:
- RF-EMF levels in pediatric units are generally low, but higher in NCCU compared to NICU.
- Exposure is likely due to wireless communication devices used by staff and parents, not medical equipment.
- Minimizing RF-EMF exposure for preterm neonates requires limiting wireless device use by staff and visitors.
Abstract:
Preterm neonates constitute a vulnerable population that is highly sensitive to its environment. Given the increased use of wireless communication devices (mobile and digital enhanced cordless telecommunications, WiFi networks, etc.), neonates hospitalized in a department of pediatrics are potentially exposed to radiofrequency electromagnetic fields (RF-EMF). Strikingly, data on RF-EMF levels in pediatric units have not previously been published. The objective of the present study was thus to quantify the RF-EMF levels in a 34-bed tertiary department of pediatrics with a neonatal critical care unit (NCCU) and a neonatal intensive care unit (NICU). To this end, we used triaxle antenna dosimeters to map the RF-EMF levels in the environment and to measure spot emissions from medical devices. In a first set of experiments, RF-EMF levels at 144 points in the staff area and in the children's rooms in the NCCU and NICU were evaluated over a 24-h period. In a second set of measurements performed in a Faraday chamber, we measured the RF-EMF levels emitted by the medical devices to which neonates are potentially exposed in the department of pediatrics. The RF-EMF levels were significantly higher in the NCCU than in the NICU (p < 0.05). Although the two units did not differ significantly with regard to the average maximum values, the single greatest value recorded in the NCCU (6 V/m GSM + UMTS 900 (UL) frequency band, in the staff area) was more than twice that recorded in the NICU (3.70 V/m in the UMTS 2100 (UL) frequency band, in the children's rooms). The NCCU and NICU did not differ significantly with regard to the time during which the RF-EMF level at each measurement point was more than two standard deviations above its mean. The RF-EMF level was significantly higher during the day than during the night (p < 0.001). The various medical devices used in the NICU did not emit detectable amounts of RF. Overall, RF-EMF levels in the NCCU and NICU were very low. It is probable that the RF-EMFs measured here were primarily generated by the parents' and staff members' activities, rather than by medical devices. However, a combination of low-level, chronic exposure with transient, elevated peak values in a vulnerable population of preterm neonates may be of particular concern. In a department of pediatrics, decreasing preterm neonates' exposure to RF-EMFs should primarily involve a limitation on the use of wireless communication devices by staff members and parents.
Related Concept Videos
The Electromagnetic Spectrum
The Electromagnetic Spectrum
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Dual Nature of Electromagnetic (EM) Radiation
Wavelength is the distance between two consecutive peaks (the highest point) or troughs (the lowest point) in the wave. Frequency is the number of...
Biological Effects of Radiation
Pharmacokinetics in Pediatric Patients: Drug Excretion