A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
[Improving the quality of hematological studies using sigmametry.]
O V Petrova1,2, G R Shabanova1, O I Murygina1
1Federal state budgetary establishment «Federal center cardiovascular surgery» (Astrakhan) Russia, 414011, Astrakhan, Russia.
Insights
Sigmetry analysis of hematology studies demonstrated good quality for most parameters. This quality control method can enhance the accuracy of platelet count determination in clinical laboratories.
Area of Science:
- Clinical laboratory science
- Hematology
- Quality control
Background:
- Hematology studies are crucial for diagnosing and monitoring various medical conditions.
- Ensuring the accuracy and reliability of hematology test results is paramount for patient care.
- Sigmetry provides a robust framework for evaluating analytical performance in laboratory testing.
Purpose of the Study:
- To evaluate the implementation quality of hematology studies using sigmetry.
- To assess the performance of hematology analyzers based on sigmetry metrics.
- To determine optimal strategies for internal laboratory quality control.
Main Methods:
- Daily monitoring of key hematology parameters: white blood cells, red blood cells, platelets, and hemoglobin.
- Utilization of a certified three-level control material ('CBC-XE Hematology control') for laboratory quality control.
- Analysis of data using sigmetry, calculating coefficient of variation (CV), systematic error (Bias), and total allowable analytical error (TE max).
- Testing performed on 'Sysmex XT-2000i' and 'Sysmex XT-1800i' hematology analyzers.
Main Results:
- On the 'Sysmex XT-2000i' analyzer, sigmetry values for platelets ranged between 3 and 4.
- For white blood cells, red blood cells, and hemoglobin on the 'Sysmex XT-2000i', sigmetry values exceeded 4.
- On the 'Sysmex XT-1800i' analyzer, sigmetry values for platelets, white blood cells, red blood cells, and hemoglobin were all greater than 4.
- The results provide a basis for selecting appropriate control materials and frequency for internal quality control.
Conclusions:
- Sigmetry analysis indicates acceptable to excellent quality for most hematology parameters tested.
- The implementation of sigmetry in daily laboratory activities can improve the quality of platelet count determination.
- Further observation is recommended to fully validate the impact of sigmetry on platelet count accuracy.
Abstract:
The purpose - evaluated the quality of the implementation of hematology studies using sigmametry. Studies were perfomed at the Federal state budgetary establishment "Federal center cardiovascular surgery". The data were collected in the period for 2017 y. The following hematology parameters were monitored daily: white cells, red blood cell, platelet, hemoglobin. For conducting laboratory quality control we used certified three-level (Level 1 (low) , Level 2 (normal)and Level 3 (High)) control material "CBC-XE Hematology control" ("R and D Systems", USA). The study of blood hematology parameters was carried out on an automatic hematology analyzer "Sysmex XT- 2000i" and "Sysmex XT- 1800i" ("Sysmex Corporation", Japan). Sigmametry values were calculated the CV (coefficient of variation), the systematic error (Bias) and the total allowable analytical error (TE max). On the analyzer «Sysmex XT- 2000i» sigma values for platelet were between 3 and 4. For white blood cells, red blood cells, hemoglobin the value of the sigma was greater than 4. On the analyzer «Sysmex XT- 1800i» for platelet, white blood cells, red blood cells, hemoglobin the value of the sigma was greater than 4. Based on the results obtained, we drew conclusions on the choice of control material and the frequency of the internal laboratory quality control. The introduction of sigmametry in the daily activities of clinic laboratory will improve the quality of the determination of platelet, which requires further observation.