Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
[Immunochromatographic test for determination of D-dimer.]
Y A Akinshina1, A V Nikitina1, E A Amelina1
1CJSC "EKOlab", 142530, Elektrogorsk, Moscow region.
Abstract:
The developing and the testing results of an immunochromatographic test for the D-dimer qualitative determination were presents in the article. The test was approved blood plasma samples in comparison with a quantitative enzyme-linked immunosorbent assay (ELISA). The results of assay with developed test were the same with ELISA results for 87,1% and 100% for samples with increased (more than 400 ng/ml FEU) and normal concentration of D-dimer, respectively. The immunochromatographic test for determination of D-dimer can be included in the diagnostic strategy as a cut test after the assessment of venous thromboembolism risk.