Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Differences in Transient Thermal Response of Commercial Esophageal Temperature Probes: Insights From an Experimental
Mohit K Turagam1, Steve Miller2, Sharan Prakash Sharma3
1Icahn School of Medicine at Mount Sinai, New York, New York.
Objectives:
The purpose of this study was to evaluate the differences in transient thermal response (TTR) among various types of commercial esophageal temperature probes (ETPs) in the United States in an experimental model.
Background:
There is little information regarding the variation in TTR among various commercial ETPs that are approved for atrial fibrillation ablation.
Methods:
We compared various thermodynamic characteristics including, mean thermal time constant (τ), time to rise 1°C (T1°C), time to peak temperature (Tpeak), and decay time among 22 different ETPs. Each probe was submerged in a constant-temperature water bath maintained at 37 ± 0.5°C and then quickly (<0.5 s) submerged into another water bath at 45 ± 0.5°C. The experiments were repeated 3 times with each probe. TTR properties were compared on the basis of probe size, design, and number of sensors.
Results:
The τ was significantly higher with the larger 24- and 18-F ETPs compared with the smaller 9-F ETPs. Compared with the 18-F probe, T1°C (11.9 s vs. 5 s), Tpeak (40.3 s vs. 14.4 s), and Tdecay (92.4 s vs. 32.4 s) was shorter with the 9-F ETPs. Solid-shaft ETPs had shorter τ (8.6 s vs. 20.5 s), T1°C (4.4 s vs. 10.1 s) and Tpeak (13.5 s vs. 32.5 s) compared with acoustascopes. Multisensor ETPs had shorter τs (3.9 s vs. 9.1 s), T1°C (2.3 s vs. 5 s), and Tpeak (6.2 s vs. 14.4 s) compared with single-sensor ETPs.
Conclusions:
There is a significant variation in TTR among the various commercially available ETPs. The use of certain ETPs might result in underestimation of luminal esophageal temperature, which can potentially lead to adverse events.
More Related Videos
04:34Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...