Medium-term clinical outcomes following Xen45 device implantation
Aitor Fernández-García1, Ying Zhou1, Mercedes García-Alonso2
1IOA Madrid, C/Galileo 104, 28003, Madrid, Spain.
Purpose:
To evaluate medium-term clinical outcomes with microstent XEN® 45 Gel Stent (Allergan Dublin, Ireland) for treatment of primary open-angle glaucoma (POAG).
Material And Methods:
This is a retrospective, descriptive and observational study involving 93 eyes from 63 patients who had undergone POAG surgery with a XEN® 45 Gel Stent implantation and had been followed up and controlled between 12 and 36 months.
Results:
IOP dropped from 18.23 ± 5.00 mmHg pre-op to 14.16 ± 2.14, 14.47 ± 2.16 and 14.63 ± 1.91 at 1, 2 and 3 years after surgery (p = 0.000, 0.000 and 0.001) consecutively. Mean number of medications dropped from 1.87 ± 0.94 preoperatively to 0.31 ± 0.69, 0.34 ± 0.63 and 1.00 ± 0.88 (p = 0.000, 0.000 and 0.017) at 12, 24 and 36 months. Mean visual field deviation values never turned out to be significant for any of the follow-up visit data. A total of 94.6% of the surgical procedures turned out to be complication-free. In one surgery, the procedure failed and 18 months later other device was implanted.
Conclusion:
POAG surgical procedures with XEN® 45 Gel Stent implants are a safe and effective treatment approach.