Background:
Polysomnography is the gold standard for detection of central sleep apnea in patients with stable heart failure. However, this procedure is costly, time consuming, and a burden to the patient and therefore unsuitable as a screening method. An electronic health (eHealth) app to measure overnight oximetry may be an acceptable screening alternative, as it can be automatically analyzed and is less burdensome to patients.
Objective:
This study aimed to assess whether overnight pulse oximetry using a smartphone-compatible oximeter can be used to detect central sleep apnea in a population with stable heart failure.
Methods:
A total of 26 patients with stable heart failure underwent one night of both a polygraph examination and overnight saturation using a smartphone-compatible oximeter. The primary endpoint was agreement between the oxygen desaturation index (ODI) above or below 15 on the smartphone-compatible oximeter and the diagnosis of the polygraph.
Results:
The median age of patients was 66.4 (interquartile range, 62-71) years and 92% were men. The median body mass index was 27.1 (interquartile range, 24.4-30.8) kg/m2. Two patients were excluded due to incomplete data, and two other patients were excluded because they could not use a smartphone. Seven patients had central sleep apnea, and 6 patients had obstructive sleep apnea. Of the 7 (of 22, 32%) patients with central sleep apnea that were included in the analysis, 3 (13%) had an ODI≥15. Of all patients without central sleep apnea, 8 (36%) had an ODI<15. The McNemar test yielded a P value of .55.
Conclusions:
Oxygen desaturation measured by this smartphone-compatible oximeter is a weak predictor of central sleep apnea in patients with stable heart failure.
More Related Videos
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...