A Novel Convolutional Neural Network-Based Algorithm for Heart Rate Measurement From Ballistocardiography Signals in
Kumar Chokalingam1, Muthukumarasamy Saravanan1, Ashish Kaushal1
1Department of Clinical Research, Turtle Shell Technologies Private Limited, Ground and Mezzanine Floor, Nomads Daily Huddle, City Centre, 40, Chinmaya Mission Hospital Road, Stage 2, Hoysala Nagar, Indiranagar, Bengaluru, 560038, India, 91 9008349922.
This study introduces a new noncontact heart rate (HR) monitoring method using ballistocardiography (BCG) and artificial intelligence. The developed algorithm accurately measures HR, improving patient comfort and enabling early detection of health issues.
Area of Science:
- Biomedical Engineering
- Artificial Intelligence in Healthcare
- Physiological Monitoring
Background:
- Continuous vital sign monitoring is crucial for early detection and improved patient outcomes.
- Traditional heart rate (HR) monitoring often requires skin contact, causing patient discomfort.
- Ballistocardiography (BCG) offers a promising noncontact alternative for continuous monitoring.
Purpose of the Study:
- Develop and validate a novel algorithm for accurate, noncontact, continuous HR monitoring using BCG signals and convolutional neural networks (CNNs).
- Enhance HR measurement accuracy across diverse healthcare settings by integrating time-domain peak detection, short-time Fourier transform, and CNN models.
- Ensure the algorithm's robustness, generalizability, and clinical applicability following FDA's Good Machine Learning Practice guidelines.
Main Methods:
- A CNN model was developed, trained on over 129,000 data points from 373 participants and tuned on over 75,000 data points from 192 participants.
- The algorithm integrates time-domain peak detection with short-time Fourier transform and CNNs for HR measurement from BCG signals.
- The algorithm was rigorously tested on over 70,000 data points from 205 participants across 5 independent studies, including ICU patients, to ensure performance in diverse conditions.
Main Results:
- The algorithm achieved a mean absolute error of under 3 bpm and a detection rate exceeding 80%.
- Bland-Altman analysis showed high accuracy with a minimal bias of 0.25 and limits of agreement within 8.59 bpm.
- Deming regression yielded a Pearson correlation coefficient of 0.97, demonstrating strong alignment with reference HR measurements.
Conclusions:
- The CNN-based algorithm provides a robust solution for contactless HR monitoring, overcoming limitations of previous methods in noise management and adaptability.
- Demonstrated accuracy in noisy, real-world clinical environments highlights its potential for widespread patient monitoring.
- The technology promises improved patient comfort and facilitates early detection of clinical deterioration.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Imaging Studies for Cardiovascular System IV: CMRI
Special considerations while measuring pulse