Accuracy and Usability of a Novel Algorithm for Detection of Irregular Pulse Using a Smartwatch Among Older Adults:
Eric Y Ding1, Dong Han2, Cody Whitcomb3
1Department of Population and Quantitative Health Sciences, University of Massachusetts Medical School, Worcester, MA, United States.
Background:
Atrial fibrillation (AF) is often paroxysmal and minimally symptomatic, hindering its diagnosis. Smartwatches may enhance AF care by facilitating long-term, noninvasive monitoring.
Objective:
This study aimed to examine the accuracy and usability of arrhythmia discrimination using a smartwatch.
Methods:
A total of 40 adults presenting to a cardiology clinic wore a smartwatch and Holter monitor and performed scripted movements to simulate activities of daily living (ADLs). Participants' clinical and sociodemographic characteristics were abstracted from medical records. Participants completed a questionnaire assessing different domains of the device's usability. Pulse recordings were analyzed blindly using a real-time realizable algorithm and compared with gold-standard Holter monitoring.
Results:
The average age of participants was 71 (SD 8) years; most participants had AF risk factors and 23% (9/39) were in AF. About half of the participants owned smartphones, but none owned smartwatches. Participants wore the smartwatch for 42 (SD 14) min while generating motion noise to simulate ADLs. The algorithm determined 53 of the 314 30-second noise-free pulse segments as consistent with AF. Compared with the gold standard, the algorithm demonstrated excellent sensitivity (98.2%), specificity (98.1%), and accuracy (98.1%) for identifying irregular pulse. Two-thirds of participants considered the smartwatch highly usable. Younger age and prior cardioversion were associated with greater overall comfort and comfort with data privacy with using a smartwatch for rhythm monitoring, respectively.
Conclusions:
A real-time realizable algorithm analyzing smartwatch pulse recordings demonstrated high accuracy for identifying pulse irregularities among older participants. Despite advanced age, lack of smartwatch familiarity, and high burden of comorbidities, participants found the smartwatch to be highly acceptable.
Related Concept Videos
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse
The pulse serves as a clinical...