Related Experiment Video
Updated: Jan 2, 2026

A Fast and Quantitative Method for Post-translational Modification and Variant Enabled Mapping of Peptides to Genomes
Published on: May 22, 2018
Optimization of TripleTOF spectral simulation and library searching for confident localization of phosphorylation
Ayano Takai1, Tomoya Tsubosaka1, Yasuhiro Hirano1
1Faculty of Pharmacy, Meijo University, Nagoya, Japan.
Abstract:
Tandem mass spectrometry (MS/MS) has been used in analysis of proteins and their post-translational modifications. A recently developed data analysis method, which simulates MS/MS spectra of phosphopeptides and performs spectral library searching using SpectraST, facilitates confident localization of phosphorylation sites. However, its performance has been evaluated only on MS/MS spectra acquired using Orbitrap HCD mass spectrometers so far. In this study, we have investigated whether this approach would be applicable to another type of mass spectrometers, and optimized the simulation and search conditions to achieve sensitive and confident site localization. Synthetic phosphopeptides and enriched K562 cell phosphopeptides were analyzed using a TripleTOF 6600 mass spectrometer before and after enzymatic dephosphorylation. Dephosphorylated peptides identified by X!Tandem database searching were subjected to spectral simulation of all possible single phosphorylations using SimPhospho software. Phosphopeptides were identified and localized by SpectraST searching against a library of the simulated spectra. Although no synthetic phosphopeptide was localized at 1% false localization rate under the previous conditions, optimization of the spectral simulation and search conditions for the TripleTOF datasets achieved the localization and improved the sensitivity. Furthermore, the optimized conditions enabled sensitive localization of K562 phosphopeptides at 1% false discovery and localization rates. These results suggest that accurate phosphopeptide simulation of TripleTOF MS/MS spectra is possible and the simulated spectral libraries can be used in SpectraST searching for confident localization of phosphorylation sites.
Related Concept Videos
Phosphorylation
During phosphorylation, protein kinases transfer the terminal phosphate group of ATP to specific amino acid side chains of substrate proteins. Serine, threonine, and tyrosine are the most commonly...
Protein Kinases and Phosphatases
Protein kinases
Many proteins in the cell are regulated by phosphorylation, the addition of a phosphate group. A family of enzymes called kinases...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...