Prospective validation study of an epilepsy seizure risk system for outpatient evaluation
Sharon Chiang1, Daniel M Goldenholz2, Robert Moss3
1Department of Neurology and Weill Institute for Neurosciences, University of California, San Francisco, San Francisco, California.
Objective:
We conducted clinical testing of an automated Bayesian machine learning algorithm (Epilepsy Seizure Assessment Tool [EpiSAT]) for outpatient seizure risk assessment using seizure counting data, and validated performance against specialized epilepsy clinician experts.
Methods:
We conducted a prospective longitudinal study of EpiSAT performance against 24 specialized clinician experts at three tertiary referral epilepsy centers in the United States. Accuracy, interrater reliability, and intra-rater reliability of EpiSAT for correctly identifying changes in seizure risk (improvements, worsening, or no change) were evaluated using 120 seizures from four synthetic seizure diaries (seizure risk known) and 120 seizures from four real seizure diaries (seizure risk unknown). The proportion of observed agreement between EpiSAT and clinicians was evaluated to assess compatibility of EpiSAT with clinical decision patterns by epilepsy experts.
Results:
EpiSAT exhibited substantial observed agreement (75.4%) with clinicians for assessing seizure risk. The mean accuracy of epilepsy providers for correctly assessing seizure risk was 74.7%. EpiSAT accurately identified seizure risk in 87.5% of seizure diary entries, corresponding to a significant improvement of 17.4% (P = .002). Clinicians exhibited low-to-moderate interrater reliability for seizure risk assessment (Krippendorff's α = 0.46) with good intrarater reliability across a 4- to 12-week evaluation period (Scott's π = 0.89).
Significance:
These results validate the ability of EpiSAT to yield objective clinical recommendations on seizure risk which follow decision patterns similar to those from specialized epilepsy providers, but with improved accuracy and reproducibility. This algorithm may serve as a useful clinical decision support system for quantitative analysis of clinical seizure frequency in clinical epilepsy practice.
More Related Videos
07:31Implementation of a Real-Time Psychosis Risk Detection and Alerting System Based on Electronic Health Records using CogStack
Published on: May 15, 2020
09:16Use of a Wireless Video-EEG System to Monitor Epileptiform Discharges Following Lateral Fluid-Percussion Induced Traumatic Brain Injury
Published on: June 21, 2019
Related Concept Videos
Epilepsy and Seizures: Overview
Various factors can trigger epilepsy, including genetic factors, brain damage, metabolic causes, and unknown etiology. Diagnosis of epilepsy involves electroencephalography (EEG), which...
Seizures: Classification
Seizures are typically classified into two main categories: focal and generalized seizures.
Focal Seizures
Focal seizures originate from specific regions of the brain. These seizures are further sub-classified into two types: