Yoshiyuki Amari1, Adriano Piattelli2,3, Karol Alí Apaza Alccayhuaman4
1ARDEC Academy, Ariminum Odontologica, Viale Giovanni Pascoli 67, 47923, Rimini, Italy.
International Journal of Implant Dentistry
|December 6, 2019
Summary
Dental implant insertion torque impacts healing. While higher torque (~30 Ncm) showed a trend toward better osseointegration, very low torque (<10 Ncm) implants can still achieve good bone integration.
Area of Science:
- Biomaterials science
- Dental implantology
- Histomorphometry
Background:
- Dental implant stability is crucial for osseointegration.
- Insertion torque is a key factor influencing primary implant stability.
- Understanding the impact of varying insertion torques on bone healing is essential for clinical success.
Purpose of the Study:
- To histomorphometrically evaluate and compare bone healing around dental implants placed at standard (~30 Ncm) versus very low (<10 Ncm) insertion torque values.
- To assess the new bone formation and bone density in relation to implant surface at different torque levels.
- To determine if very low insertion torque compromises implant osseointegration.
Main Methods:
- Twelve patients received two screw-shaped titanium implants in the mandible, one at <10 Ncm and one at ~30 Ncm insertion torque.
- Implants were allowed to heal in a non-submerged manner for 8 weeks.
- Histological analysis of retrieved biopsies was performed to quantify new bone contact and density.
Main Results:
- New bone in contact with the implant surface was 39.3% at <10 Ncm and 49.4% at ~30 Ncm.
- Total mineralized bone was 46.8% at <10 Ncm and 57.0% at ~30 Ncm.
- No statistically significant differences were found in new bone or total mineralized bone density between the two groups.
Conclusions:
- A trend towards higher osseointegration was observed with ~30 Ncm insertion torque compared to <10 Ncm.
- Implants placed with very low insertion torque (<10 Ncm) can achieve successful osseointegration.
- Clinical outcomes suggest that achieving good implant integration is possible even with minimal initial torque.