Validation of Waste Anaesthetic Gas Exposure Limits When Using a Closed Vaporizer Filling System: A Laboratory-Based
Shane Varughese1, H Peter Bacher2
1Global Medical Affairs, AbbVie Inc, North Chicago, IL, USA. shane.varughese@abbvie.com.
Introduction:
It is desirable to minimise exposure of personnel to halogenated inhaled anaesthetics in the operating room to avoid deleterious short-term and long-term health effects. The objective of this study was to determine whether, while filling anaesthetic vaporizers with sevoflurane using AbbVie's closed vaporizer filling system (Quik-Fil™), concentrations of sevoflurane in ambient air remained at or below recommended levels when measured at different operator heights.
Methods:
Nine filling runs were conducted, with measurement heights of 95, 130, 140, 150, 160, and 185 cm. Within each 15-min run, five vaporizers were sequentially filled from bottles of sevoflurane with the closed valving system. Ambient-air sevoflurane concentration in the breathing zone was continuously measured once per second by using a MIRAN SapphIRe 205BXL portable ambient air analyser.
Results:
The use of the closed filling system maintained a level of waste anaesthetic gas exposure that was well below (mean, 0.10 ppm; maximum, 0.16 ppm) the recommended short-term value of 20 ppm average for 15 min provided by the Swedish Work Environment Authority and also fell below the US limit of a time-weighted average of 2 ppm provided by the National Institute for Occupational Safety and Health. Exposure to sevoflurane appeared to be independent of the height at which the measurement was made.
Conclusions:
The presence of sevoflurane in the work environment while using the closed filling system maintains a level of waste anaesthetic gas exposure well below the recommended levels at all tested operator heights.