Stress distribution of complete-arch implant-supported prostheses reinforced with silica-nylon mesh
Tarcisio-José de A Paes-Junior1, João-Paulo-Mendes Tribst2, Amanda-Maria-de Oliveira Dal Piva2
1DDs, MSc, PhD, Adjunct Professor, Department of Dental Materials and Proshodontics, São Paulo State University (Unesp), Institute of Science and Technology, São José dos Campos / SP, Brazil.
Background:
This study evaluated the presence of a silica-nylon mesh and two cantilever lengths on the biomechanical behavior of complete-arch implant-supported prostheses.
Material And Methods:
Twenty-four (24) complete mandibular arch implant-supported prostheses were divided into 4 groups according to the presence of reinforcing mesh (with or without) and the cantilever length (molar - 15 mm or premolar - 5 mm). The specimens were submitted to strain gauge analysis (30-kgf, 10 s) at different points (molar and premolar). Three-dimensional models were created based on the in vitro specimens, and the results in the bone (microstrain), prostheses (tensile stress), implants and prosthetic screws (von-Mises stress) were evaluated using the finite element method (FEM). All materials were considered homogeneous, isotropic and linear. Strain gauge data were submitted to 3-way analysis of variance and the Tukey test (α=.05). FEM results were qualitatively analyzed using colorimetric graphs.
Results:
The microstrain magnitude for the prostheses with reinforcement was 519.91±359 and 583.33±661 without reinforcement (p=.001). The microstrain values for loading on the molar was 867.49±784 and on the premolar was 235.75±145. FEM corroborated with the in vitro findings for the bone behavior. The load application in the premolar showed reduced stress concentration, and a significant difference was observed between the presence or absence of the reinforcement for the prostheses.
Conclusions:
Silica-nylon mesh reduced the peri-implant microstrain and the prosthesis stress regardless of the cantilever extension. For temporary complete-arch implant-supported prostheses, the limitation of the cantilever to the premolar region improves the biomechanical response during load application. Key words:Finite element analysis, biomechanical response, dental implants, prosthetic dentistry.