Clinical effects of calibration of CoaguChek XS PT Test Strips on Anticoagulation Management Services
Insights
The CoaguChek XS PT Test Strip recalibration and recall did not significantly impact patient care in an anticoagulation clinic. Most elevated INR results were managed without serious clinical events.
Area of Science:
- Medical Devices
- Clinical Diagnostics
- Hematology
Background:
- Roche Diagnostics recalibrated CoaguChek XS PT Test Strips in January 2018.
- An urgent correction statement and Class I recall were issued due to inaccurate International Normalized Ratio (INR) results > 4.5.
Purpose of the Study:
- To assess the clinical impact of the CoaguChek XS PT Test Strip recalibration and recall on patients managed by Anticoagulation Management Services (AMS).
Main Methods:
- Two retrospective chart reviews were conducted for AMS patients before and after the recall announcement.
- Data included INR values, confirmatory testing, reasons for elevation, and interventions.
Main Results:
- A small percentage of patients experienced elevated INRs (>4.5) attributed to the recalibration.
- Most elevations were managed by holding medication doses, with no reported bleeding events or hospitalizations.
Conclusions:
- The CoaguChek XS PT Test Strip calibration update and subsequent recall had a minimal clinical impact on this anticoagulation clinic's patient population.
- Confirmatory testing revealed discrepancies but did not lead to adverse patient outcomes.
Objective:
In January 2018, Roche Diagnostics calibrated CoaguChek XS PT Test Strips to the latest World Health Organization reference preparation. On September 12, 2018, Roche Diagnostics issued an urgent medical device correction statement warning of inaccurate international normalized ratio (INR) results greater than 4.5 and recommended additional confirmatory testing. Eight weeks later, on November 2, the U.S. Food and Drug Administration issued a class I recall of all affected lot numbers. The objective of this project was to determine the clinical effects of recalibration of CoaguChek on patients managed by the Anticoagulation Management Services (AMS).
Methods:
Two retrospective chart reviews were conducted for patients seen by AMS from May 1 to September 11 and from September 12 to November 2, 2018. Data collected included date of INR, confirmatory testing (if applicable), reason for elevated INR, and interventions. Data were analyzed using descriptive statistics.
Results:
From May 1 to September 11, 2018, 16 of 166 patients were affected by recalibrated strips, resulting in 19 of 1176 INRs being greater than 4.5. Almost half (47%) of elevations were due to unknown causes, and most were corrected by holding at least 1 dose. Of the 152 patients treated from September 12 to November 2, 8 patients were affected by the recalibration, resulting in 11 of 453 INRs being greater than 4.5. Six resulted in confirmatory INR testing with an average difference of 2 points. No bleeding events or hospitalizations were reported in either group.
Conclusion:
Based on the experiences of 1 anticoagulation clinic, the CoaguChek XS PT Test Strip calibration update and class I recall did not result in a meaningful clinical impact.
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Venous Thrombosis III: Interprofessional Care
Errors occurring during blood pressure monitoring
Several factors...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Disorders of Hemostasis
Thromboembolic Disorders
Two factors primarily cause thromboembolic conditions.