Evaluation of the performance of the Cepheid Xpert HIV-1 Viral Load Assay for quantitative and diagnostic uses
Laura Wesolowski1, William Fowler1, Wei Luo1
1Division of HIV/AIDS Prevention, National Center for HIV/AIDS, Viral Hepatitis, STD and TB Prevention, Centers for Disease Control and Prevention, Atlanta, GA, USA.
Background:
Cepheid's Xpert HIV-1 Viral Load (Xpert VL), a simplified, automated, single-use quantitative assay used with the GeneXpert System, is not FDA approved.
Objectives:
Using stored plasma, we conducted a study to assess the ability of Xpert VL to quantify viral load relative to the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 (Cobas VL) and to examine the use of the Xpert VL as a qualitative diagnostic test.
Study Design:
Following HIV-1 viral stock dilutions, we conducted a probit analysis to identify the concentration where 95 % of specimens had quantified VLs. We also examined Xpert and Cobas log VL correlation in linearity panels; compared the proportion of 220 seroconverter specimens with virus detected using McNemar's test; and tested specimens from persons with untreated, established HIV-1 infection (n=149) and uninfected persons (n=497). Furthermore, we examined Xpert VL as a qualitative test in seroconverter specimens with early (n=20) and later (n=68) acute infections.
Results:
At 1.80 log10 copies/mL, 95 % of specimens had quantifiable virus using Xpert VL. Xpert and Cobas VLs were highly correlated (R2=0.994). The proportion of seroconverter specimens with virus detected using Cobas and with Xpert VL was not statistically different (p=0.0578). Xpert VL detected 97.9 % of established infections, and specificity was 99.80 % (95 % CI 98.87%-99.99%). Xpert VL detected 90 % and 98.5 % of early and later acute infections, respectively.
Conclusions:
If approved, Xpert VL could allow U.S. laboratories that cannot bring on large, complex testing platforms to conduct HIV monitoring. An approval for diagnostic use may provide timely identification of HIV infections.