Related Experiment Video
Updated: Jan 1, 2026

Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021
In Vivo Evaluation of 3 Electronic Apex Locators: Root ZX Mini, Apex ID, and Propex Pixi
Gabriela Serna-Peña1, Susana Gomes-Azevedo1, Jorge Flores-Treviño1
1Department of Endodontics, Universidad Autónoma de Nuevo Léon, Monterrey, Nuevo Léon, Mexico.
Introduction:
The aim of this in vivo study was to compare the accuracy of 3 electronic apex locators (EALs) (Root ZX mini [J Morita Corp, Tokyo, Japan], Apex ID [SybronEndo, Glendora, CA], and Propex Pixi [Dentsply Maillefer, Ballaigues, Switzerland]) to determine the working length.
Methods:
Thirty single-rooted human teeth that were scheduled for extraction were selected for the study. Electronic measurements were performed with the 3 EALs. After the teeth had been extracted, a #10 K-file was used to determine the actual working length, which was established at 0.5 mm short of the major foramen. The data were statistically analyzed with analysis of variance (α = 0.05).
Results:
No significant differences were found among the experimental groups (P > .05). The mean distance from the actual working length to the file tip was 0.163 ± 0.032 mm when Root ZX mini was used, 0.343 ± 0.032 mm for Propex Pixi , and 0.012 ± 0.008 mm for Apex ID.
Conclusions:
Under the in vivo conditions of this study, no statistically significant differences were observed among the 3 EALs.