High incidence of diaphragmatic myopotential oversensing by a specific implantable cardioverter defibrillator
Samuel H Baldinger1, Andreas Haeberlin1, Helge Servatius1
1Department of Cardiology, Inselspital, Bern University Hospital and the University of Bern, Bern, Switzerland.
Introduction:
Diaphragmatic myopotential oversensing (dMPO) by implantable cardioverter defibrillators (ICDs) is thought to be a rare condition that can be misdiagnosed as lead failure and lead to unnecessary lead replacement. We observed several cases of dMPO in patients with Sorin/LivaNova ICDs (MicroPort Sci.). We sought to systematically assess the incidence of dMPO in patients with Sorin/LivaNova ICDs.
Methods And Results:
A predefined number of 100 consecutive patients with Sorin/LivaNova ICDs were prospectively included in the device clinic of our center. Stored arrhythmia episodes were checked for spontaneous dMPO. In addition, we performed provocation maneuvers by Valsalva. At least one episode of spontaneous or provoked dMPO was seen in 12 (12%) of the 100 patients included in the study (86% males, median age: 66 years). Nine of 89 patients (10%) with true bipolar and 3 of 11 patients (27%) with integrated bipolar sensing configuration were affected. Spontaneous dMPO was observed in 7 of 58 patients (12%) with sensitivity programmed to 0.4 mV and in 2 of 42 patients (5%) with sensitivity programmed to 0.6 mV (not significant). In three patients, dMPO could be provoked with no spontaneous episodes recorded. In two nonpacemaker-dependent patients with a CRT-D, ventricular pacing was temporarily inhibited. No antitachycardia therapy was triggered by dMPO in any patient.
Conclusions:
DMPO is frequent in patients with Sorin/LivaNova ICDs, especially with sensitivity programmed to 0.4 mV. It also frequently occurs with true bipolar sensing configuration. DMPO should not be misinterpreted as lead failure to avoid unnecessary lead replacement.
More Related Videos
Related Concept Videos
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Mechanism of Cardiac Arrhythmias
Cardiomyopathy III: Hypertrophic Cardiomyopathy