Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Comparison of image quality between synthetic and patients' electrocardiogram-gated 320-row pediatric cardiac
Eriko Maeda1, Go Shirota2, Eisuke Shibata3
1Department of Radiology, Graduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo, 113-8655, Japan. NQF38509@nifty.com.
Background:
Pediatric cardiac computed tomography (CT) can be acquired without electrode placement by using synthetic electrocardiogram (ECG).
Objective:
To determine whether the depiction of gross cardiac structures and coronary arteries in 320-row pediatric CT is not inferior when CT is gated with synthetic ECG at 150 beats per minute (bpm), compared to the patients' own ECG.
Materials And Methods:
Sixty 320-row CT examinations performed in children younger than 3 years old with congenital cardiac anomaly were enrolled in this retrospective study. Thirty examinations were scanned using the children's own ECG for gating and 30 examinations were scanned using synthetic ECG at 150 bpm. The image quality was compared between the two gating modes using a 3-point scale to delineate the following anatomical structures: atrial septum, ventricular septum, right atrium, right ventricle, left atrium, left ventricle, main pulmonary artery, ascending aorta, aortic arch including the patent ductus arteriosus, descending aorta, right coronary artery and left main trunk. Beam-hardening artifacts from contrast enhancement material were evaluated using a 3-point scale, and the overall image quality was evaluated using a 5-point scale.
Results:
Synthetic ECG was not inferior to the patients' ECG in depicting each structure, beam-hardening artifact and overall image quality. Average indices were clinically acceptable imaging quality, except for subjective image quality of mid and distal coronary arteries.
Conclusion:
Pediatric cardiac CT in patients younger than 3 years old can be acquired using synthetic ECG gating, with image quality not inferior to the patients' ECG.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT