Contamination of ionization chambers by talcum powder
W F Hanson1, D J Arnold, R J Shalek
1Accredited Dosimetry Calibration Laboratory, University of Texas M. D., Anderson Cancer Center, Houston 77030.
Abstract:
This is a word of caution to anyone using ionization chambers protected by thin rubber sheaths in water. Four Farmer-type ionization chambers contaminated with talcum powder were received for calibration by the Accredited Dosimetry Calibration Laboratory at the University of Texas M. D. Anderson Cancer Center. The chambers show a marked energy dependence (5% to 20%) to soft orthovoltage x rays. The response of the contaminated chambers is compared with the chambers' response before contamination and after cleaning. Techniques for identifying contaminated chambers and suggestions for cleaning them are presented.
More Related Videos
07:54Experimental Methods of Dust Charging and Mobilization on Surfaces with Exposure to Ultraviolet Radiation or Plasmas
Published on: April 3, 2018
09:29An Air-liquid Interface Bronchial Epithelial Model for Realistic, Repeated Inhalation Exposure to Airborne Particles for Toxicity Testing
Published on: May 13, 2020
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...